• <tr id='9DIrbD'><strong id='9DIrbD'></strong><small id='9DIrbD'></small><button id='9DIrbD'></button><li id='9DIrbD'><noscript id='9DIrbD'><big id='9DIrbD'></big><dt id='9DIrbD'></dt></noscript></li></tr><ol id='9DIrbD'><option id='9DIrbD'><table id='9DIrbD'><blockquote id='9DIrbD'><tbody id='9DIrbD'></tbody></blockquote></table></option></ol><u id='9DIrbD'></u><kbd id='9DIrbD'><kbd id='9DIrbD'></kbd></kbd>

    <code id='9DIrbD'><strong id='9DIrbD'></strong></code>

    <fieldset id='9DIrbD'></fieldset>
          <span id='9DIrbD'></span>

              <ins id='9DIrbD'></ins>
              <acronym id='9DIrbD'><em id='9DIrbD'></em><td id='9DIrbD'><div id='9DIrbD'></div></td></acronym><address id='9DIrbD'><big id='9DIrbD'><big id='9DIrbD'></big><legend id='9DIrbD'></legend></big></address>

              <i id='9DIrbD'><div id='9DIrbD'><ins id='9DIrbD'></ins></div></i>
              <i id='9DIrbD'></i>
            1. <dl id='9DIrbD'></dl>
              1. <blockquote id='9DIrbD'><q id='9DIrbD'><noscript id='9DIrbD'></noscript><dt id='9DIrbD'></dt></q></blockquote><noframes id='9DIrbD'><i id='9DIrbD'></i>

                卧式蒸汽灭菌器⌒的维护保养

                卧式蒸汽灭菌器操作系统■采用先进的微机自动处理技术,灭菌运行过ㄨ程实行微机自动控制,操控台上的液晶显示窗显示整个灭菌动态和功能运行情况,产品主要部件采用不锈钢材料精制而成,灭菌指标达〇到国家医用手术级标准,是各医疗卫生、大专院校、食品化工、生物科研等单位对器械、敷料、食品器皿、培养基等物品进行消毒灭菌的理想产品。
                卧式蒸◥汽灭菌器的维护保养:
                1.每天进行灭菌器门、仪表的表面擦拭,灭菌间地面清洁々至少一次。
                2.每天清理灭菌器内排泄口处滤网的杂质,避免灭菌器运行中杂质进入排气管。
                3.每天运行前检查灭♀菌器门封是否平整、完好,无脱出和破损。
                4.每天应检查仪表指针的准确度,观察灭∮菌器运行停止后,温度仪表、仪表指针是否归在“0”位;观察蒸汽、水等介质管路和阀件有无泄露;观察灭∮菌器运行指示灯是否完好;一旦发现以上部件出现问题,不应使用灭菌№器,经维护修理后使用。
                5.每周进行灭菌器内的清洁擦拭,彻底擦╱拭清理。
                6.每季度进行灭菌设备外部的清洁,避免积尘,缩短空气滤器的使用寿命。应避免元器件与水接触,一旦湿水应擦干后方可接通电源。
                7.每@ 季度检查各连线的插座、接头是否松动,松动的应插紧。
                8.每6个〓月清理安全阀表面。
                9.每周检查清理蒸汽管路过滤器一次,记录结果。
                10.每年灭菌器进行年≡检一次,安全阀、表、温度表每年效验至少一次,检查结果记录并留存。空气过滤▲器应定期更换。